Abstract

In this study, we report the development and validation of an LC-tandem mass spectrometry method for the simultaneous quantitation of bendamustine and copanlisib in mouse plasma as per the US FDA regulatory guidelines. The sample processing involves extraction of bendamustine and copanlisib along with internal standard (IS; warfarin) from 50 μL mouse plasma using a liquid-liquid extraction method. The chromatographic separation of bendamustine, copanlisib and the IS was achieved on an Atlantis dC18 column using an isocratic mobile phase (5 mM ammonium acetate:methanol, 20:80 v/v). Bendamustine, copanlisib and the IS eluted at 0.88, 1.39 and 0.74 min, respectively, with a total run time of 2.5 min. The calibration curve ranged from 3.99-2996 and 4.33-3248 ng/mL for bendamustine and copanlisib, respectively. Inter- and intra-day precision and accuracy, stability in processed samples and upon storage, dilution integrity and incurred sample reanalysis were investigated for both the analytes. The intra- and inter-day precisions were in the ranges of 2.01%-5.05% and 2.74%-6.13% and 1.98%-7.64 and 8.62%-9.04% for bendamustine and copanlisib, respectively. Stability studies showed that both analytes were stable on bench top for 6 h, in auto-sampler for 24 and at -80°C for 30 days. The validated method was successfully applied to a pharmacokinetic study in mice.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.