Aims: In this study, a simple, green, and rapid capillary zone electrophoresis (CZE) method coupled with a diode array detector (DAD) was applied for the analysis of avanafil (AVA) and dapoxetine hydrochloride (DAP) as a binary mixture using vardenafil (VAR) as an internal standard (IS) in pure form and pharmaceutical formulation.
 Methodology: The separation was done using fused silica capillary (58.5 cm total length, 50 cm effective length, and 50 μm internal diameter) and the running background electrolyte (BGE) was 100 mM acetate buffer at pH 3.6. During the separation process, the applied voltage was 30 KV, while the temperature was 25 °C. The sample injection was applied at a pressure of 50 mbar for 10 s, and detection was carried out at 210 nm for DAP and 248 nm for AVA and VAR.
 Results: Analysis of the tested drugs and the internal standard was carried out in less than 6.5 min, where the migration times were 4.29, 4.90, and 6.02 min for IS, DAP and AVA respectively. The proposed method showed linearity in the concentration range 5-80 and 5-70 μg/mL with correlation coefficients 0.9996 and 0.9999 for AVA and DAP respectively. The limit of detection (LOD) was 0.523 and 0.531 for AVA and DAP respectively, while the limit of quantification (LOQ) was 1.585 and 1.608 in respective order. The Peak purity and identity in the proposed method were validated by DAD.
 Conclusion: The proposed CZE method was validated according to ICH guidelines and applied successfully for the estimation of AVA and DAP in their combined pharmaceutical preparation.

Highlights

  • AVA is a class of drugs called phosphodiesterase (PDE) inhibitors, and it works by improving blood flow to the penis during sexual stimulation leading to an erection [1]

  • In this study, a simple, green, and rapid capillary zone electrophoresis (CZE) method coupled with a diode array detector (DAD) was applied for the analysis of avanafil (AVA) and dapoxetine hydrochloride (DAP) as a binary mixture using vardenafil (VAR) as an internal standard (IS) in pure form and pharmaceutical formulation

  • The proposed CZE method was validated according to International Conference on Harmonization (ICH) guidelines and applied successfully for the estimation of AVA and DAP in their combined pharmaceutical preparation

Read more

Summary

Results

Analysis of the tested drugs and the internal standard was carried out in less than 6.5 min, where the migration times were 4.29, 4.90, and 6.02 min for IS, DAP and AVA respectively. The proposed method showed linearity in the concentration range 5-80 and 5-70 μg/mL with correlation coefficients 0.9996 and 0.9999 for AVA and DAP respectively. The limit of detection (LOD) was 0.523 and 0.531 for AVA and DAP respectively, while the limit of quantification (LOQ) was 1.585 and 1.608 in respective order. The Peak purity and identity in the proposed method were validated by DAD. Conclusion: The proposed CZE method was validated according to ICH guidelines and applied successfully for the estimation of AVA and DAP in their combined pharmaceutical preparation

INTRODUCTION
CE System
Materials and Chemicals
General Procedure
Capillary conditioning
Preparation of standard solutions and construction of calibration graphs
Assay of tablets dosage form
Buffer type and Ph
Buffer concentration
Sample injection and use of internal standard
Selection of detection wavelength
Linearity and ranges
Limit of detection and quantification
Accuracy and precision
Selectivity
Robustness
Stability of solutions
Application of Dosage Forms