Abstract

A novel diffusion cell has been constructed which allows study of membrane diffusion processes without the need for sampling of the receiver compartment, that is highly sensitive and, being based around a diode array spectrophotometer also allows for continuous, real-time recording of multi-species concentration changes in the receiving compartment. The system is controlled to operate isothermally (via a Peltier control system) at temperatures between 15 and 85 °C. To examine the performance of this novel design, the transfer of tetracaine from a preparation in PEG 400 (20% tetracaine in PEG 400) has been studied. The results have been used to determine flux, lag time and related parameters. The performance of the novel cell is compared with results from traditional Franz cell diffusion studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.