Abstract
Possibilities to use the MAXIPIX 2×2 detector for coherent scattering applications are investigated. This study focuses on X-ray Photon Correlation Spectroscopy experiments. Flatfield measurements show that the MAXIPIX detector pixels respond very inhomogeneously in the X-ray range between 5 and 10 keV most commonly used for coherent scattering experiments. Two colloidal model sample systems are investigated using the MAXIPIX and a direct illumination CCD for comparison. It is demonstrated that the MAXIPIX 2×2 detector is applicable for static and dynamic coherent scattering experiments after flatfield correction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.