Abstract

A high-resolution magnetic sector mass spectrometer has been developed and used for transient kinetic measurements on reactions that are important for automotive exhaust gas catalysis. This instrument allows separation of the isobaric masses of CO and N2 as well as CO2 and N2O, which are separately monitored in time at a sampling rate of 48 Hz. The time resolution is sufficient to determine individual kinetic rate parameters for reactions under transient conditions. A description of the experimental set-up as well as the software used is given. The results are illustrated by the catalytic reduction of NO by CO. To the best knowledge of the authors, this is the first application of a magnetic sector mass spectrometer for kinetic research under dynamic operation conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.