Abstract

Extrapulmonary tuberculosis (EPTB) remains a challenging diagnosis. The purpose of this study was to assess the accuracy of Xpert MTB/RIF Ultra (Cepheid, USA) for rapid diagnosis of EPTB in Tunisia. Eight hundred and forty-seven extrapulmonary samples collected from 2017 to 2021, were subjected to Xpert MTB/RIF Ultra. Microscopy and culture were performed for all the specimens. The accuracy of Xpert Ultra was evaluated in comparison to the culture. Xpert Ultra diagnosed EPTB with a global sensitivity of 80.66% (74.3–85.75) and specificity of 70.87% (67.31–74.20). The molecular test was most accurate when performed in cerebrospinal fluids, bones and joints and cutaneous specimens showing a sensitivity of 100% and a specificity ranging from 70.60 to 91.11%. In lymph node samples comprising aspirates and biopsies, the sensitivity of Xpert Ultra was high 87.50% (77.23–93.53), however, the specificity was 51.08% (44.67–57.46). For pleural samples, the Xpert Ultra sensitivity was 77.50% (68.34–84.68) ranging from 71.43 to 80% in pleural biopsies and fluids respectively. The specificity in all pleural specimens was 79.56% (74.40–83.91). Xpert Ultra showed promise in the diagnosis of EPTB. The performances varied according to the site of the disease. The test may be more valuable if used in combination with other diagnostic modalities.