Abstract

To report the usefulness of intraoperative real-time adjustment of intraocular lens (IOL) tilt during the intrascleral fixation with intraoperative optical coherence tomography (iOCT) as a clinical evaluation and investigate the factors contributing to IOL tilt using iOCT as an experimental evaluation. Retrospective cohort study and experimental research. As a clinical evaluation, the medical records of 43 eyes of 41 patients who underwent intrascleral IOL fixation combined with real-time iOCT observation were retrospectively reviewed. As an experimental evaluation, in order to investigate the factors contributing to IOL tilt, the four experiments were performed using iOCT. The mean IOL tilt angle (°) at the end of surgery and 3 months after surgery were 1.81 ± 1.15 and 2.10 ± 1.66, respectively (p = 0.46). No apparent intra- or postoperative complications occurred during the follow-up period. The experimental evaluation indicated that the IOL tilt was influenced by the insertion angle of the haptic in the vertical direction. The mean IOL tilt angle (°) was 1.94 ± 0.09, 4.67 ± 0.11, 8.90 ± 0.11, and 15.78 ± 0.85 when the insertion angle of the haptic was 0°, 10°, 27.5°, and 45° in the vertical direction, respectively (p < 0.01). Clinical and experimental IOL tilt assessment using iOCT is interactively useful for better quality surgery and better postoperative outcome.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call