Abstract
Several drug substances or excipients are hygroscopic. The uptake or loss of water of such substances is generally difficult to control during processing or storage of drug products. DSC instruments with sub-ambient temperature equipment allow the determination of the amount of freezable water by measuring the corresponding melting enthalpy. The determination of freezable water adds valuable information complementary to TG analysis for understanding the processing and storage of raw materials and drug products. Several substances were tested as is, without treatment, after storage at 92% r.h. and after equilibration with water. The results of these experiments showed that it was possible to demonstrate defined hydrate formation, to determine the upper level of binding of water in amorphous substances and to confirm reversible hydrate formations demonstrated by temperature resolved X-ray diffraction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.