Abstract
A toroidal grating monochromator (TGM) based beamline was designed earlier and it is being setup on the 450 MeV Indus-1 synchrotron radiation source. This beamline operates in 40–1000 A wavelength range and gives a moderate resolution (∼500) at high throughput. Now-a-days, spherical gratings are giving very high resolution in some complex monochromators. We have evaluated the use of spherical gratings in the simple rotation mount of a TGM. The motivation has been to evaluate if the existing TGM beamline, without any hardware changes, can be used with the much cheaper spherical gratings. The two monochromators, employing toroidal and spherical gratings respectively, have been compared by performing detailed ray tracing calculations. Our study shows that, with the same mechanical hardware, a real spherical grating monochromator gives better performance compared to the conventional TGM with respect to the spectral resolution and the flux. Furthermore, for moderate resolution this improvement is realized even when the exit slit is not moved to keep the spherical grating in-focus.