Abstract

Solid-phase extraction with two-step elution has been developed for effective elimination of copper and iron interference with mercury determination by flow-injection cold vapour atomic absorption spectrometry (CV AAS). Sodium tetrahydroborate(III) was used as reducing agent. Cation-exchanger Dowex 50Wx4 was applied for the sorption of mercury and both interfering ions. In the first step elution of Cu(II) and Fe(III) was performed using 0.5 mol L(-1) KF solution. Then mercury was eluted with 0.1% thiourea in 8% HCl. The detection limit (3 delta) for Hg(II) was 27 ng L(-1). The expanded uncertainty estimated for the whole procedure was about 6%. The accuracy of the proposed method was evaluated by determination of the recovery of known amount of mercury added to mineral, spring, and tap waters, and by analysis of a certified reference material BCR-144R (sewage sludge).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.