Abstract
Several decomposition procedures and their influence on the determination of mercury by electrothermal (ET) and cold vapour (CV) atomic absorption spectrometry (AAS) have been studied. Soil samples were decomposed by microwave digestion in closed and open vessels as well as by digestion under reflux according to German standard. The use of different acids (HNO3, HCl or aqua regia) was evaluated and compared in respect to their influence on the determination of mercury by ET AAS and CV AAS. The digestion solutions were analyzed by ET AAS with a palladium modifier and by CV AAS using SnCl2 or NaBH4, as reducing agents. The detection limits obtained with different procedures were also evaluated. For the soil containing 6.25 μg/g of Hg the ET AAS measurements were possible. In the case of lower concentration of mercury the CV AAS determination following the microwave digestion procedure with HCl or aqua regia is recommended. The accuracy of the proposed procedure was confirmed by the determination of total mercury in SRM 2711 Montana Soil.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.