Abstract

Universal screening for anemia is important in children, but invasive blood sampling is required. A new device (Radical-7® Pulse CO-Oximeter™, Masimo, Irvine, CA, USA) now enables non-invasive hemoglobin concentration (SpHb) measurement to be done, but the usefulness of this device for anemia screening in children is unclear. The objective of this study was to compare SpHb with complete blood count (CBC) using a hematology analyzer (Microsemi® LC-667CRP; Fukuda Denshi, Tokyo, Japan). SpHb measurement with Radical-7® was done as part of a medical check-up in 3-year-old children (n = 110). Another 43 pediatric patients were checked for CBC using Microsemi® and monitored with Radical-7®. The mean SpHb level of the 3-year-old children was 12.1 ± 0.64 g/dL (range, 10.8-13.7 g/dL). The correlation of Radical-7® and Microsemi® was 0.602 (P < 0.0001). On Bland-Altman comparison, bias was -0.6 ± 1.1 g/dL. Even though further improvement is required, Radical-7® offers many possibilities in the context of primary screening.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.