Abstract

The effects of intravenous injection of indigo carmine on noninvasive and continuous total hemoglobin (SpHb) measurement were retrospectively evaluated with the Revision L sensor. The subjects were 18 patients who underwent elective gynecologic surgery under general anesthesia. During surgery, 5mL of 0.4% indigo carmine was injected intravenously, and changes in SpHb concentrations between before and after the injection were evaluated. The mean age was 52.4±12.8years. Before injection, the median SpHb level was 10.1 (range, 6.8-13.4)g/dL. The results demonstrated no change in SpHb concentration between before and after indigo carmine injection as detected by the Revision L sensor. SpHb measurements as determined with the Revision L sensor were not affected, even after the intravenous injection of indigo carmine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.