Abstract Active personal dosemeters (APD) have been found to be very efficient tools to reduce occupational doses in many applications of ionizing radiation. In order to be used in interventional radiology and cardiology (IR/IC), APDs should be able to measure low energy photons and pulsed radiation with relatively high instantaneous personal dose equivalent rates. A study concerning the optimization of the use of APDs in IR/IC was performed in the framework of the ORAMED project, a Collaborative Project (2008–2011) supported by the European Commission within its 7th Framework Program. In particular, eight commercial APDs were tested in continuous and pulsed X-ray fields delivered by calibration laboratories in order to evaluate their performances. Most of APDs provide a response in pulsed mode more or less affected by the personal dose equivalent rate, which means they could be used in routine monitoring provided that correction factors are introduced. These results emphasize the importance of adding tests in pulsed mode in type-test procedures for APDs. Some general recommendations are proposed in the end of this paper for the selection and use of APDs at IR/IC workplaces.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.