Abstract

ObjectiveTo investigate the consistency of the recommended methodology, we analyzed the dosimetric results obtained for each setup beam shaping filter and tube voltage and evaluated the uncertainty associated with the full dose measurement process. MethodsA 300-mm PTIC was used to validate the results obtained with the 100-mm PTIC. Considering the other dosimetric parts, a cylindrical 160-mm diameter PMMA phantom and a cylindrical 320-mm diameter PMMA phantom were also used in the experimental protocol. ResultsFor the lowest and highest tube voltages available, the CTDIair,160 values obtained by 1-step dosimetry with the 300-mm PTIC were greater than the respective values obtained by 2- and 3-step dosimetry with the 100-mm PTIC. ConclusionThis study established that careful positioning of the 100-mm PTIC in 2 or 3 steps, as well as proper execution of the other dosimetric parts recommended by the IAEA, represents a validated approach within up to 20% uncertainty for wide cone beam CT dosimetry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.