Abstract

A silicon-intensified target vidicon tube (SIT) is evaluated for analytical use in the determination of 17 elements in an Ar-separated C2H2/air flame and 5 elements in an Ar-separated C2H2/N2O flame using combined atomic emission and atomic fluorescence as excited by a 150 W CW EIMAC xenon arc lamp. Benefits of combining emission and fluorescence signals in multielement analysis are experimentally shown. Limits of detection are approximately 100 times worse than previously reported by atomic fluorescence alone using a conventional spectrophotometer with the same 150 W EIMAC lamp. Linear dynamic ranges are between ∼10 and ∼100 for elements analyzed in the C2H2/air flame and are as high as 1000 for elements in the C2H2/N2O flame. Little or no use of the SIT and many similar image detectors for multielement analysis via atomic emission and/or atomic fluorescence is predicted at the present “state of the art.”

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.