Abstract

Implantable cardioverter defibrillators (ICDs) are frequently offered to patients with arrhythmogenic right ventricular dysplasia/cardiomyopathy (ARVD/C). Yet ICDs in these patients may be complicated by poor sensed amplitudes resulting from fatty and fibrous tissue replacement of right ventricular myocardium. We present the case of a patient with ARVD/C who had inappropriate detection of ventricular tachycardia with a single-chamber ICD due to poor sensed right ventricular amplitudes. We discuss how the use of a bipolar coronary sinus lead and a biventricular ICD generator with a novel header configuration solved the problem.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.