Abstract

A chemiluminometric method for the automated flow injection analysis of ammonia is described. The essence of the invention is the use of a bioreactor consisting of both immobilized L-glutamate dehydrogenase (GLDH) and L-glutamate oxidase (GLXD), which are sequentially aligned in this order in a minicolumn measuring 2.0 X 20 mm. The unidirectional constant flow of liquid through the column reactor minimizes the reversed diffusion of the solutes so that the following sequence of reactions is ensured. Thus, ammonia to be determined is first transformed by GLDH into L-glutamate, which then produces hydrogen peroxide by GLXD. Hydrogen peroxide in the effluent from the column is then determined by its chemiluminescence upon admixing with luminol and potassium ferricyanide. The present method gives linearity of the standard curve for ammonia up to 1.0 mM. It is at least 100 times more sensitive than the conventional method for ammonia assay using ultraviolet absorption measurement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.