Abstract

In pyroprocessing, uranium (U) is recovered from molten LiCl–KCl salt, and, for safeguard purposes, it is important to analyze the U and Plutonium (Pu) concentrations in a timely manner. In the present work, salt samples containing U were fabricated. The laser used in the present work was an Nd:YAG laser with a wavelength of 532 nm, a laser energy on the sample of 11.5 mJ, and a pulse repetition rate of 10 Hz. The plasma emission light was measured with an Echelle spectrometer. A total of 100 points on the sample surface were measured as the laser incident position was changed. The U and potassium (K) peaks in the spectrum were identified. Univariate and multivariate analyzes were conducted to determine the accuracy and limit of detection (LOD) of the laser-induced breakdown spectroscopy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.