Abstract

Ultra performance liquid chromatography–mass spectrometry (UPLC–MS) method has been developed for the simultaneous determination of pharmaceuticals in industrial waste water samples. The binary mobile phase was used a mixture of acetonitrile and 0.1% aqueous solution of formic acid at a flow rate of 0.4mlmin−1. Identification and quantitation of the analyzed drug samples were carried out using mass spectrometer equipped with electrospray ionization source operated in negative ionization mode. The linear range of the proposed method was found to be in the range of 18–270ngml−1 for all drugs with limit of detection in the range of 1.25 to 1.89ngml−1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.