Abstract

The aim of this study was to describe the indications for upgrade of pacemakers (PMs) or single chamber (VVIR) ICDs to dual chamber (DDDR) ICDs, surgical approach, hardware hybridization, and clinical outcome. Patients with preexisting PMs or VVIR ICDs may develop indications for ICD therapy or dual chamber pacing, respectively, that can be served by DDDR ICDs that incorporate preexisting transvenous leads. Fifty-seven patients underwent upgrade from PMs (29/57) or VVIR ICDs (28/57) to pectoral DDDR ICDs. Preexisting transvenous atrial and/or ventricular leads suitable for continued use were incorporated into new DDDR ICDs in 88.5% and 100% of PM and VVIR ICD upgrades, respectively. Acceptable DFTs were achieved in 56 (98.2%) of 57 patients. Appropriate VT/VF therapies were registered among 33.3% of patients during follow-up. No shocks due to lead noise were observed in any patient with hybridized transvenous leads. Atrial far-field R wave (FFRW) oversensing occurred in 24% of DDDR ICD systems incorporating a preexisting atrial lead. FFRW was overcome by programming reduced atrial sensitivity without interfering with the normal ICD system performance in all instances. Upgrade of PMs and VVIR ICDs to pectoral DDDR ICDs is safe and technically feasible in most patients. Preexisting transvenous leads can be successfully incorporated into new DDDR ICDs, simplifying the surgical procedure, minimizing transvenous hardware, and eliminating the possibility of hazardous pacemaker-ICD interactions.