Abstract
Improvement of a conventional JEOL Raman spectrometer with a single channel photo multiplier detector is described. New optical components [fibers, mirror, lens, and charge coupled device detector (CCD)] have been chosen to design a high quality and easy‐to‐use instrument. Tests have shown that with this modified spectrometer, Raman spectra can be acquired of a quality comparable to the spectra obtained previously, but the time needed to obtain a spectrum is markedly reduced. Selected test spectra and a simple calibration procedure to obtain the wavenumber values from the band CCD pixel position are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.