Abstract
Magnetic resonance imaging (MRI) artifacts of adjustable shunt devices are thought to be similar to metal clip artifacts, in that they are larger with higher field strength scanners. We have published several reports about the artifacts of new MRI-resistant adjustable shunt devices, and we found a case in which a 3.0-T scanner showed smaller artifacts than the 1.5-T scanner. We aimed to clarify whether this claim is true or not. Under permission of our institutional Ethical Committee, 2 volunteers underwent imaging studies using 3.0-T and 1.5-T scanners from GE, Siemens, and Philips. Four MRI-resistant adjustable shunt devices-proGAV2.0 (Miethke), Codman Certas Plus (Johnson & Johnson), Polaris (Sophysa), and Strata MR valve (Medtronic)-were fixed on the left temporal scalp. Routine MRI images, including T1-and T2-weighted imaging, fluid-attenuated inversion recovery, diffusion-weighted imaging (DWI), and magnetic resonance angiography (MRA), were obtained. We also compared artifacts between a 3.0-T scanner and a-1.5 T scanner in 4 patients. The 3.0 T-scanners showed smaller artifacts than the 1.5-T scanners on DWI and MRA images for all shunt devices and scanners. In the other sequences, the results depended on the MRI scanner manufacturer; however, the GE 3.0-T scanner showed smaller artifacts in every sequence. This was also true in the 4 clinical cases. A 3.0-T scanner is recommended over a 1.5-T scanner for patients with MRI-resistant adjustable shunt devices in the diagnosis of acute ischemic condition or when using GE scanners.