Abstract

Background: Line-edge roughness (LER) is often measured from top-down critical dimension scanning electron microscope (CD-SEM) images. The true three-dimensional roughness profile of the sidewall is typically ignored in such analyses. Aim: We study the response of a CD-SEM to sidewall roughness (SWR) by simulation. Approach: We generate random rough lines and spaces, where the SWR is modeled by a known power spectral density. We then obtain corresponding CD-SEM images using a Monte Carlo electron scattering simulator. We find the measured LER from these images and compare it to the known input roughness. Results: For isolated lines, the SEM measures the outermost extrusion of the rough sidewall. The result is that the measured LER is up to a factor of 2 less than the true on-wafer roughness. The effect can be modeled by making a top-down projection of the rough edge. Our model for isolated lines works fairly well for a dense grating of lines and spaces as long as the trench width exceeds the line height. Conclusions: In order to obtain and compare accurate LER values, the projection effect of SWR needs to be taken into account.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.