Abstract

One of the authors previously developed an unconstrained and noninvasive measurement system of heartbeat and respiration using an acoustic sensor enclosed in an air pillow or air mat. The system measured instantaneous heartbeat and respiration periods by applying Kalman filters and maximum likelihood method to the dynamic models of the heartbeat and respiration signals extracted from the sensor output. To improve the measurement accuracy, this paper proposes a more stable and accurate measurement system of heartbeat and respiration taking into account not only the harmonic component of the heartbeat, but also the sinusoidal component of the frequency of the sum of the fundamental mode frequencies of the heartbeat and respiration signals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.