Abstract

We present an integrating hemisphere-based (i.e.,a variant of integrating spheres) implementation of the indirect illumination method for absolute photoluminescence quantum yield measurements, which is a recommended method in the international standard IEC 62607-3-1:2014. We rigorously formulated a mathematical model and a measurement procedure for the absolute photoluminescence quantum yield measurement in the integrating hemisphere-based system. The measurement system was calibrated using an Hg-Ar discharge lamp and spectral irradiance standard lamps for wavelength and relative spectral radiant flux scales, respectively. Furthermore, we identified and evaluated uncertainty components involved in the photoluminescence quantum yield (PLQY) measurement. To validate our measurement system, we applied it to the two de facto standard dyes: quinine bisulfate (QBS) and fluorescein (FLS). Consequently, their PLQY values were determined to be 0.563±0.024 (k=2) and 0.876±0.032 (k=2) for, respectively, QBS and FLS, which are consistent with previous reports.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.