Abstract

This paper describes a detailed derivation of the uncertainty budget for measurements of the absolute spectral radiance factor with PTB’s gonioreflectometers. The measurement uncertainty is determined according to the ‘Guide to the expression of uncertainty in measurement’ provided by the Bureau International des Poids et Mesures. To give an insight, the uncertainty analysis is applied to measurements of three ceramic samples. The influence of different systematic and statistical sources of uncertainty and their contributions are discussed in detail. For wavelengths larger than 450 nm the total measurement uncertainty is dominated by systematic contributions. Towards UVA the major contribution is caused by statistical effects, which are mainly due to the decreasing power of the radiation source used. We present an approach capable to reduce this contribution for the VIS/UVA-transition spectral range by means of a specially designed LED-based sphere radiation source.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.