Abstract

This paper deals with an uncertainty analysis of gas flow measurements using a compact, high-speed, clearance-sealed realization of a piston prover. A detailed methodology for the uncertainty analysis, covering the components due to the gas density, dimensional and time measurements, the leakage flow, the density correction factor and the repeatability, is presented. The paper also deals with the selection of the isothermal and adiabatic measurement models, the treatment of the leakage flow and discusses the need for averaging multiple consecutive readings of the piston prover. The analysis is prepared for the flow range (50 000:1) covered by the three interchangeable flow cells. The results show that using the adiabatic measurement model and averaging the multiple readings, the estimated expanded measurement uncertainty of the gas mass flow rate is less than 0.15% in the flow range above 0.012 g min−1, whereas it increases for lower mass flow rates due to the leakage flow related effects. At the upper end of the measuring range, using the adiabatic instead of the isothermal measurement model, as well as averaging multiple readings, proves important.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.