Abstract

Abstract Upgrades to the vacuum-wavelength calibration service at the National Institute of Standards and Technology are reported. The instrumentation centerpiece is an optical frequency comb stabilized to a GPS-disciplined oscillator, thereby providing direct traceability to the SI second. Historically, the service has covered lasers at the popular interferometry wavelengths red and green. Recently, capability has been added for calibrating wavemeters at multiple telecom wavelengths in the range (1520 < \lambda < 1570) nm. For most commercially-available wavemeters, the test-uncertainty-ratio is about 10^4.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.