Abstract

Ultra-trace analysis of actinides, such as Pu isotopes and 237Np, in bioassay samples is often needed for radiation protection programs at nuclear facilities. Accelerator mass spectrometry (AMS), particularly the compact ETH Zurich system “Tandy”, has evolved over the years as one of the most sensitive, selective, and robust techniques for actinide analysis. Employment of the AMS technique can reduce the demands on sample preparation chemistry and increase sample analysis throughput, due to very low instrumental detection limit, high rejection of interferences, and low susceptibility to adverse sample matrices. Initial research and development tests were performed to explore and demonstrate the analytical capability of AMS for Pu and Np urine bioassay. In this study, urine samples spiked with femtogram levels of Np and Pu isotopes were prepared and measured using compact ETH AMS system and the results showed excellent analytical capability for measuring Np and Pu isotopes at femtogram/litre levels in urine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.