Abstract

This study aimed to investigate the effectiveness and safety of modified thread carpal tunnel release (mTCTR) using Smartwire-01 in patients with carpal tunnel syndrome (CTS). Patients with CTS who required CTR were enrolled. Symptom severity and functional status were assessed using the Boston Carpal Tunnel Syndrome Questionnaire-Symptom Severity Scale (BCTQ-SSS) and Functional Status Scale (BCTQ-FSS), and pain was assessed using a numerical rating scale (NRS) at 4, 8, and 12 weeks after mTCTR. The scores were compared with the pre-procedural scores. The electrophysiologic study and median nerve cross-sectional area (CSA) measurements at the wrist before and 12 weeks after mTCTR were compared. A total of 11 patients were included. No adverse effects were reported throughout the study period. The NRS, BCTQ-SSS, and BCTQ-FSS scores significantly improved at 4 weeks after mTCTR, and this improvement persisted throughout the follow-up period (NRS and BCTQ-SSS, P < 0.001; BCTQ-FSS, P = 0.012). After 12 weeks, the latency and velocity of the median sensory nerve action potential significantly improved, compared with those before mTCTR (latency, 5.4 ± 1.3 to 4.7 ± 1.1 ms, P = 0.01; velocity 27.8 ± 6.8 to 31.8 ± 7.4 m/s, P = 0.019). No significant change was observed in the median nerve CSA before and after mTCTR. mTCTR using Smartwire-01 is a safe and effective procedure and a possible alternative to surgery.