Abstract

The Ar-N2 mix gas effect can easily improve the sensitivity of ICPMS; however, this effect discriminates against Sr. In this study, it was found that Ar-N2 mixed gases introduced into nebulizing gas enhanced the sensitivity of online solid-phase extraction (SPE) with inductively coupled plasma mass spectrometry (ICPMS) for radioactive strontium quantification. An ultrasonic nebulizer (USN) improved the Ar-N2 mixture gases effect of Sr and the mix gases (with USN) enhanced 3.7-times the signal intensity of Sr in normal pure Ar gas (with USN) in an online SPE-ICPMS. By adapting the gas-loading means from a nebulizing gas unit via USN, no careful tuning was necessary for the plasma turning. With this signal enhancement, a 0.06 pg/L detection limit (0.3 Bq/L) was achieved for radioactive strontium (90Sr) in online SPE-ICPMS within 30 min. In addition, environmental paddle water in the Fukushima Nuclear Power Plant was measured and the valued correspond to that obtained by radiometry.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call