Abstract

Internal standardization is employed to compensate for ionization suppression in inductively coupled plasma-mass spectrometry (ICP-MS). By examination of the response of over 50 elements to a sodium matrix under different operating conditions, it is apparent that an internal standard is most effective when it is close in mass and ionization energy to the analyte. The extent of suppression and the relative order of suppression of various analyte elements can differ for various matrix elements. Generally, precision was improved by the use of internal standardization; the extent of improvement differed for different analyte elements and operating conditions. A comparison between ICP-MS with ultrasonic and pneumatic nebulization is described. The ultrasonic nebulizer usually exhibits better sensitivity and detection limits for analyte elements, unless the extent of suppression induced by the concomitant matrix is very high.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.