Abstract

We characterized a vertically configured electrical detection system that used scanning tunneling microscopy (STM) to detect antigen-antibody binding. This technique could be used to easily construct a multiple measurement system in a protein chip. We utilized immunocomplexes comprised of our model protein, prostate specific antigen (PSA), corresponding antibody fragments, and gold nanoparticle-antibody conjugates. The electrical tunneling current between the STM tip and these complexes exhibited a peaklike pulse, the frequency of which depended on the surface density of the bound complexes. We could therefore quantitatively measure PSA concentrations as low as 10fg∕mL using periodogram analysis of this peak frequency.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.