A pressure-controlled headspace solid-phase microextraction (PC-HS-SPME) setup was developed, by reconsidering the strengths and weaknesses points of the similar reported systems. The new setup was coupled with gas chromatography–flame ionization detection (GC–FID) for direct analysis of benzene, toluene, ethylbenzene and xylene (BTEX) in contaminated soils, without any sample preparation step. The important experimental factors, affecting the performance of the method, including volumes of extraction and vacuum vials, type of SPME fiber, extraction time and temperature, moisture content of the sample, and sonication time were studied and optimized. Under the optimal conditions, good linearity of the calibration curves (R2 > 0.997) was obtained in the concentration range of 0.1–20,000 ng g−1. The limits of detections were found to be 0.001–0.08 ng g−1. The relative standard deviations, for six repetitive analyses of 100 ng g−1 BTEX, were obtained to be 5.7–12.3%. The PC-HS-SPME–GC–FID procedure was successfully applied for the extraction and determination of BTEX in the polluted soil samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.