Abstract

A high-sensitive cholesterol amperometric biosensor based on the immobilization of cholesterol oxidase (ChOx) onto the ZnO nanoparticles has been fabricated which shows a very high and reproducible sensitivity of 23.7 μA mM −1 cm −2, detection limit (based on S/N ratio) 0.37 ± 0.02 nM, response time less than 5 s, linear range from 1.0 to 500.0 nM and correlation coefficient of R = 0.9975. A relatively low value of enzyme’s kinetic parameter (Michaelis–Menten constant) ∼4.7 mM has been obtained which indicates the enhanced enzymatic affinity of ChOx to Cholesterol. To the best of our knowledge, this is the first report in which such a very high-sensitivity and low detection limit has been achieved for the cholesterol biosensor by using ZnO nanostructures modified electrodes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.