Abstract

This paper reports the fabrication of highly-sensitive cholesterol biosensor based on cholesterol oxidase (ChO x) immobilization on well-crystallized flower-shaped ZnO structures composed of perfectly hexagonal-shaped ZnO nanorods grown by low-temperature simple solution process. The fabricated cholesterol biosensors reported a very high and reproducible sensitivity of 61.7 μA μM −1 cm −2 with a response time less than 5 s and detection limit (based on S/N ratio) of 0.012 μM. The biosensor exhibited a linear dynamic range from 1.0–15.0 μM and correlation coefficient of R = 0.9979. A lower value of apparent Michaelis–Menten constant ( K m app), of 2.57 mM, exhibited a high affinity between the cholesterol and ChO x immobilized on flower-shaped ZnO structures. Moreover, the effect of pH on ChO x activity on the ZnO modified electrode has also been studied in the range of 5.0–9.0 which exhibited a best enzymatic activity at the pH range of 6.8–7.6. To the best of our knowledge, this is the first report in which such a very high-sensitivity and low detection limit has been achieved for the cholesterol biosensor by using ZnO nanostructures modified electrodes.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call