Abstract

Ammonia nitrogen is a common contaminant in water and its determination is important for environmental protection. In this paper, an electrochemical sensor based on an ultramicro interdigitated array electrode (UIAE) chip with optimized construction was fabricated with Micro-Electro-Mechanical System (MEMS) technology and developed to realize the detection of ammonia nitrogen in water. The effects of spacing-to-width ratio and width of the working electrode on UIAE's electrochemical characteristics and its ammonia nitrogen detection performance were studied by finite element simulation and experiment. The results demonstrated that the smaller the spacing-to-width ratio, the stronger generation-collection effect, and the smaller the electrode width, the stronger the edge effect, which led to an easier steady-state reach, a higher response current, and better ammonia nitrogen determination performance. The fabricated UIAE chip with optimized construction showed the linear detection range of 0.15 mg/L~2.0 mg/L (calculated as N), the sensitivity of 0.4181 μA·L·mg-1, and good anti-interference performance, as well as a long lifetime. UIAE based on bare Pt was successfully applied to ammonia nitrogen detection in water by optimizing structure, which might broaden the methods of ammonia nitrogen detection in water.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call