Abstract

A very quick and easy LC-MS/MS analysis method for 5-HIAA (5-hydoxyindoleacetic acid) has been developed. The method was fully validated and proved to work well in a clinical setting. Precision at the upper reference limit 123 nmol/L was 3,3% CV. Accuracy ranged from 96% at low levels (50–100 nmol/L) to 99.7% at high levels (500 nmol/L). A previously reported reference interval of 35–123 nmol/L was confirmed as valid based on analysis of 40 samples from voluntary blood donors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.