Abstract

Our research group has developed a thermoluminescence dosimeter (TLD) based on pellets of LiF:Mg,Ti mixed with polytetrafluoroethylene (LiF:Mg,Ti +PTFE). This TLD can be used as a personal dosimeter. Extensive type testing, carried out with reference to the International Electrotechnical Commission (IEC) Standard, were performed for the purpose of accepting the LiF:Mg,Ti+PTFE as a personal TL dosimeter. Tests performed include repeatability, batch homogeneity, linearity, detection threshold, and light sensitivity. Results showed that locally made LiF:Mg,Ti+PTFE TLDs met all the standard requirements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.