Abstract

Abstract In order to replace LiF commercial thermoluminescence dosemeters (TLDs) now used in personnel monitoring, we developed a TLD personnel monitoring badge based on pellets of CaSO4:Dy bound in pure KBr placed in plastic holders provided with 0.8 mm Pb filters to compensate for the energy dependence of CaSO4:Dy. Performance tests, based on the proposals of international standards for TLDs intended to be used as general deep tissue personal dosemeters, were done by developing the detailed test methods both for the tests required by the International Atomic Energy Agency (fading, linearity, effect of fluorescent light and sunlight and stability under standard climatic condition) and to meet the requirements of the International Electrotechnical Commission (batch homogeneity, detection threshold, stability of zero point repeatibility and residual). Initial tests showed a high residual and an increase of inhomogeneity. However, these drawbacks were overcome by changing the annealing procedure and the readout conditions. Using these new conditions TLDs met all the requirements for personal dosemeters.