Abstract

Because of their inert character and desired biocompatibility, titanium implants have been universally accepted as safer alternatives to the conventional stainless steel orthopedic implants; however, recent emergence of type IV hypersensitivity reactions to titanium have included eczema, contact dermatitis, a prolonged febrile state, sterile osteonecrosis, and impaired fracture and wound healing. This report presents a patient with postoperative incision dehiscence and devascularization of surfaces in contact with titanium hardware after undergoing a double calcaneal osteotomy and a first metatarsal-cuneiform arthrodesis using titanium alloy implants. Titanium hypersensitivity was confirmed in this case through standard allergy patch testing by a board-certified immunologist. Complete healing occurred after diagnosis of the titanium allergy and hardware explant. To our knowledge, this is one of a few known allergies to titanium implants after foot and ankle surgery.