High frequency induction heating infrared absorption method is a relative measurement method, and it requires calibrating the analyzer during the measurement process. Usually, one-time calibration on an analyzer is made for high sulfur value by using standard substances. The method proposed by this paper calibrates carbon-sulfur analyzers with a method of two-time calibration at both ends, which eliminates the impact of sulfur blank and thus improves the accuracy and precision of ultra-low sulfur measurement. The method extents the lower limit of measurement of sulfur in high temperature alloy, to 0.00005%, namely, 0.5ppm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.