Abstract
Two liquid chromatography (LC) methods with fluorimetric detection have been developed to measure atenolol and propranolol in human plasma. The same 5 μm Nucleosil RP-18 column, extraction procedure and mobile phase (containing acetonitrile, water, triethylamine and phosphoric acid, pH 3) were used. The linearity ranges were 25–800 ng/ml for atenolol and 3.13–100 ng/ml for propranolol. The coefficients of variation for validation assays were lower than 15% at the concentration assayed. The functions of the analytical error were linear: SD (ng/ml)=7.698+0.037C for atenolol and SD (ng/ml)=0.126+0.036C for propranolol.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.