Abstract

The response of a large-area two-dimensional (2D) thermoluminescence (TL) dosimetry system with 20 × 20 cm2 TL LiF:Mg,Cu,P foils, developed at the Institute of Nuclear Physics PAN was studied for doses of therapeutic proton beams with energies of 80 MeV, 150 MeV and 225 MeV. The one-hit detector model fitted to measured dose response yielded characteristic doses of 227 Gy, 209 Gy and 203 Gy respectively. The system was applied to investigate geometrical parameters of spots produced by proton pencil beams at the Bronowice Cyclotron Centre IFJ PAN. Mean spot sizes measured with TL foils were compared with mean spot sizes calculated for Gafchromic® films. It was shown that the 2D TL system was capable to measure the spot size up to peak doses of 20 Gy without additional correction for dose response. For measurements of the spot size using Gafchromic® EBT3 films an additional calibration must be applied to correct for the non-linear response for doses exceeding 1 Gy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.