Abstract
Two-dimensional gel electrophoresis (2-DE) is a protein separation technique often used to separate plasma or serum proteins in an attempt to identify novel biomarkers. This protocol describes how to run 2-DE gels using narrow pH 3-5.6 immobilised pH gradient strips to separate 2 mg of serum proteins. pH 3-6 ampholytes are used to enhance the solubility of proteins in this pH range before the serum proteins are separated in the first dimension by isoelectric point (isoelectric focusing) followed by molecular weight (SDS-PAGE). This approach using the pH 3-5.6 range differs from pH ranges more commonly used for serum or plasma biomarker discovery which span three or more pH units (e.g. pH 3-10 and 4-7), and has the advantage that the pH range lies outside the range of three highly abundant proteins and therefore improves separation and representation of low abundance features. The protocol described takes approximately 8 days.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.