Abstract

A single-chord CO2 (10.6-μm) He–Ne (3.39-μm) two-color laser interferometer which has been developed for the ETA-BETA II RFP machine is presented. Compared to the usual 0.633-μm vibration compensation systems, the 3.39-μm interferometer allows faster optical path-length variations to be measured without losing fringes, can stand larger angular deflections of the beams (e.g., due to vibrations of the optic mounts) without losing alignment, and uses the same acousto-optic modulator as the CO2 system, thus minimizing the differences in the optical path of the two interferometers. The instrument has been extensively utilized on ETA-BETA II, where line density variations larger than 1020 m−2 and occurring in less than 4 μs can be routinely measured and the residual error after vibration compensation is ≊2×1018 m−2. Some further developments of the diagnostic are presented, which are presently underway or planned.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.