Abstract
Totally implanted central venous access devices for chemotherapy (chemoport) are being used increasingly in lung cancer patients. Vascular catheters are associated with various complications including infection, thrombosis as well as spontaneous fractures and embolization of the catheter, which is known as 'pinch-off syndrome'. 'Pinch-off syndrome' refers to the compression of a subclavian central venous catheter between the clavicle and first rib resulting in an intermittent or permanent obstruction, which can lead to tears, transection, or embolization. We report two cases of fractured and embolized implanted subclavian venous catheters in which the fragments were removed percutaneously. A 62-year-old man presented with back pain with a duration of a few weeks. The chest radiograph revealed complete transsection and embolization of the catheter into the right atrium. In addition, a 47-year-old woman with a chemoport had a grade 3 pinch-off sign in a chest radiograph demonstrating complete transsection and embolization of the catheter into the pulmonary artery. Both cases were managed by retrieving the embolized distal fragment percutaneously and removing the proximal section of the catheter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.