Abstract
The current study compared the effectiveness of the graduated tuning fork (128 Hz) and the neurothesiometer in assessing vibration sensation perception in patients presenting with type II diabetes mellitus. A quota sample of patients (n = 21; age range 43-73 years) were assessed using the neurothesiometer and tuning fork by two investigators at five sites on both feet. There was a positive correlation between the results for the two methods of assessment for both investigators, and also between the results for both tools at three individual sites. Overall, there was 66.2% agreement between the results obtained from the two investigators using the tuning fork at each site; however, Kappa values only reached statistical significance at one site, indicating variability between the results from the two tools. This study suggests that assessment of vibration sensation with the tuning fork may be unreliable. These preliminary findings are based on a small sample size; thus further research is warranted.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.