.Significance: Recovering accurate oxygenation estimations in the breast with quantitative photoacoustic tomography (QPAT) is not straightforward. Accurate light fluence models are required, but the unknown ground truth of the breast makes it difficult to validate them. Phantoms are often used for the validation, but most reported phantoms have a simple architecture. Fluence models developed in these simplistic objects are not accurate for application on the complex tissues of the breast.Aim: We present a sophisticated breast phantom platform for photoacoustic (PA) and ultrasound (US) imaging in general, and specifically for QPAT. The breast phantom is semi-anthropomorphic in distribution of optical and acoustic properties and contains wall-less channels with blood.Approach: 3D printing approaches are used to develop the solid 3D breast phantom from custom polyvinyl chloride plastisol formulations and additives for replicating the tissue optical and acoustic properties. A flow circuit was developed to flush the channels with bovine blood with a controlled oxygen saturation level. To showcase the phantom’s functionality, PA measurements were performed on the phantom with two oxygenation levels. Image reconstructions with and without fluence compensation from Monte Carlo simulations were analyzed for the accuracy of oxygen saturation estimations.Results: We present design aspects of the phantom, demonstrate how it is developed, and present its breast-like appearance in PA and US imaging. The oxygen saturations were estimated in two regions of interest with and without using the fluence models. The fluence compensation positively influenced the estimations in all cases and confirmed that highly accurate fluence models are required to minimize estimation errors.Conclusions: This phantom allows studies to be performed in PA in carefully controlled laboratory settings to validate approaches to recover both qualitative and quantitative features sought after in in-vivo studies. We believe that testing with phantoms of this complexity can streamline the transition of new PA technologies from the laboratory to studies in the clinic.

Highlights

  • Cancerous tissue is often associated with higher density of blood vessels than healthy tissue due to the process of angiogenesis.[1]

  • The oxygen saturations were estimated in two regions of interest with and without using the fluence models

  • We believe that testing with phantoms of this complexity can streamline the transition of new PA technologies from the laboratory to studies in the clinic

Read more

Summary

Introduction

Cancerous tissue is often associated with higher density of blood vessels than healthy tissue due to the process of angiogenesis.[1] Tumor vasculature is known to be abnormal in architecture. Function, characterized by heterogeneity, tortuousity, chaotic branching, and leaky vessel walls.[1] These characteristics are potential hallmarks to identify the cancer. Imaging techniques, such as contrast-enhanced ultrasound imaging, dynamic contrast-enhanced magnetic resonance imaging (MRI), and positron emission tomography, can visualize tumor vasculature in the clinical setting with the use of contrast agents and are seeing application in cancer diagnosis and treatment monitoring.[2]. Optical imaging techniques are attractive candidates for imaging blood vessels, since they do not require contrast agents, owing to the high optical absorption coefficient of hemoglobin (Hb) in blood.[4,5]

Objectives
Methods
Results
Discussion
Conclusion